View Larger Image
|
|
DCN-3000PS Hand-held coating thickness gauge with integrated probe | 3 Models
DCF-3000PS - Ferrous Only Model
DCN-3000PS - Non-Ferrous Only Model
DCFN-3000PS - Combination Ferrous & Non-Ferrous Model
Measuring Range: 0-1500 µm (0-60.00 mils)
With coating thickness gauges designed for measurement on ferrous metals, non-magnetic coatings such as varnish, paint, enamel, chrome, copper, zinc, etc. can be measured on steel and iron (magnetic induction method).
Coating thickness gauges designed for measurement on non-ferrous metals can measure all electrically insulated coatings such as varnish, paint, anodizing coatings on non-ferrous metals & stainless steel (eddy-current method).
The DCFN-3000PS is an exceptionally small hand-held gauge with integrated probe for fast, practical and precise coating thickness measurement.
|

 Dependable accuracy over the entire range of the gauge, not just at "calibration check" points. | Measures to 0.1 µm (0.01 mil), automatically rounding off to the nearest 0.1 mil (1 µm). | Keypad-selectable English or metric units. | Continuous reading update detects soft coatings as probe tip "sinks" into them. |
| Convient compact design | Longer lasting, driftless calibration...
Set it and forget it for long periods of time and from job to job | Very reliable operation... Consistent, repeatable measurements | Manufacturer's Certificate Included |
|

 The DCN-3000PS is supplied as a complete kit with:
- coating thickness gauge
- non-ferrous zero-plate
- 2 calibration standards
- 2 AAA batteries
- soft carrying pouch
- protective rubber shell
- mini screw driver
- instruction manual
- manufacturer´s certificate
|  |


| Calibration Standards | Calibration Foils for DCF(N)-3000 Series Instruments various thicknesses: 10µm, 30µm, 40µm, 50µm, 100µm, 200µm, 360µm, 500µm, 800µm, 1000µm and 1500µm.
|  | | | Click image for more details
|
| 3000-IRP | Integrated IrDA infrared interface Archivable up to 5 years, if dryly stored without UV exposure Speed of printing: 16 lines/sec |  | | | Click image for more details
|
| 3000-IR | For data transfer to a standard PC (RS232) or printer (RS232). The 3000-IR is self-powered by the RS-232 (serial) or USB port Meets IrDA 1.0 physical specifications |  | | | Click image for more details
|
|
|
Microns (µm) |
| 0.0 - 99.9
µm |
0.1 µm |
| 100.0 -
249.8 µm |
0.2 µm |
| 250.0 -
499.5 µm |
0.5 µm |
| 500 - 999 µm |
1 µm |
| 1.000 -
1.998 mm |
0.002 mm |
| |
|
Mils |
| 0.00 - 9.99
mils |
0.01 mils |
| 10.00 -
24.98 mils |
0.02 mils |
| 25.00 -
49.95 mils |
0.05 mils |
| 50.0 - 99.9
mils |
0.1 mils |
|


| Gauge version: |
DCN-3000PS Coating thickness on non-ferrous metals |
| Range: |
0 - 1500 µm (0 - 60.00 mils) |
| Accuracy: |
±1µm (0.04 mils) + 1% of reading |
| Resolution: |
Refer To Table Above |
| Display: |
back-light, 4-digit alphanumeric, digit height 10mm |
| Min. Measuring Area: |
5 mm x 5 mm (0.2" x 0.2") |
| Min. Curvature Radius: |
convex: 3 mm (0.12") concave: 5 mm (0.2") |
| Min. Substrate Thickness: |
Ferrous: 0.5 mm (20 mils) Non-Ferrous: 50 µm (2 mils) |
| Calibration: |
factory calibration, zero calibration, foil calibration, offset-function: addition or subtraction of a constant value |
| Memory: |
80 readings |
| Statistics Program: |
Number of readings, mean value, standard deviation, maximum and minimum reading of max. 10,000 readings |
| Set Limits: |
Adjustable selectable with acoustic alarm system |
| Data Output: |
Infrared IrDA Standard |
| Operating Temp: |
0°C to 50°C (32°F to 122°F) |
| Surface Temp: |
-15°C to 60°C (5°F to 140°F) |
| Ambient temperature: |
0 ... 50°C /32° F-122°F |
| Power supply: |
2 x micro batteries (AAA) 1.5 V |
| Dimensions: |
99mm x 48mm x 24 mm / 4 |
| Weight: |
85 g / 3 ounces(gauge with batteries) |
| Protection class: |
IP 52 (proof against dust and dripping water) |
| Standards: |
DIN, ISO, ASTM, BS |
| Warranty: |
1 year |
|
|